Fujirebio Diagnostics, Inc. recalls Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applic
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
- Products
-
- Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applic
- Company
- Fujirebio Diagnostics, Inc.
- Where it was sold
-
US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
States: Arizona, California, Florida, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Minnesota, New Jersey, Pennsylvania, Texas, Washington
- Barcodes
05414470815579- Units affected
- 117 units
- About this recall
- Product Name: Lumipulse G pTau 217 Plasma Calibrators Model/Catalog Number: 81557 Software Version: Not Applicable Product Description: Lumipulse G pTau 217 Plasma Calibrators: The pTau 217 plasma concentration of a specimen is automatically calculated from the calibration curve, which is also automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G pTau 217 Plasma Calibrators: Liquid (Frozen), 1x1.5 mL (5 Concentrations) CAL 1 0 pg/mL pTau 217 Plasma Calibrator CAL 2 0.250 pg/mL pTau 217 Plasma Calibrator CAL 3 1.000 pg/mL pTau 217 Plasma Calibrator CAL 4 5.000 pg/mL pTau 217 Plasma Calibrator CAL 5 10.000 pg/mL pTau 217 Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 300, 0.05% ProClin 950 Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio