Mc3 Inc recalls NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygena
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
- Products
-
- NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygena
- Company
- Mc3 Inc
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Units affected
- 188 units
- About this recall
- NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E