FDAMedical devicesRecalled Status: Ongoing

Mc3 Inc recalls NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygena

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
Products
  • NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygena
Company
Mc3 Inc
Where it was sold

US Nationwide distribution.

Nationwide

Units affected
188 units
About this recall
NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E