Think Surgical, Inc. recalls Assembly
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
- Products
-
- Assembly
- Company
- Think Surgical, Inc.
- Where it was sold
-
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.
States: Alabama, California, Missouri, Nevada, Ohio, Texas, Utah
- Units affected
- 16
- About this recall
- Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System