Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Think Surgical, Inc. recalls Assembly

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
Products
  • Assembly
Where it was sold

US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.

States: Alabama, California, Missouri, Nevada, Ohio, Texas, Utah

Units affected
16
About this recall
Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System