In2Bones, SAS recalls I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- This Field Action is being conducted following the identification of a batch mix-up between I.B.S Compression screws diameter 6.5mm length 75mm and I.B.S Compression screws diameter 6.5mm length 80mm. The screws of the batches concerned by the recall may have different lengths from those issued on their labels.
- Products
-
- I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm
- Company
- In2Bones, SAS
- Where it was sold
-
US Nationwide distribution in the state of Tennessee.
States: Tennessee
- Barcodes
3760225710685- Units affected
- 59 units
- About this recall
- I.B.S¿ 6.5-C Compression screw - diam 6.5mm lg 75mm,Part Number S65 ST175