Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls HDS Traction Boot II

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for the Hip Distraction System, Traction Boot II, due to the boots being assembled incorrectly. The locking ridge on the front face of the boot key socket is facing upwards, which prevents the socket mechanism on the traction boot from attaching properly to the spar of the traction system. This will cause the boot to detach from the traction system.
Products
  • HDS Traction Boot II
Where it was sold

US Nationwide distribution in the state of Florida.

States: Florida

Units affected
80 units
About this recall
HDS Traction Boot II, REF AR-6529-23, The Traction Boot is designed to position, support and/or distract the patient's foot for hip surgery