Intuitive Surgical, Inc. recalls da Vinci 5 Surgeon Console Viewer Display
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
- Products
-
- da Vinci 5 Surgeon Console Viewer Display
- Company
- Intuitive Surgical, Inc.
- Where it was sold
-
US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.
States: Arizona, California, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Mississippi, North Carolina, Nevada, New York, South Carolina, Texas, Utah, Virginia, Wisconsin
- Barcodes
00886874119747- Units affected
- 47
- About this recall
- da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45