Wilson-Cook Medical Inc. recalls Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: T
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
- Products
-
- Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: T
- Company
- Wilson-Cook Medical Inc.
- Where it was sold
-
U.S. Distribution to states of: CA, CO, IN, and NC.
States: California, Colorado, Indiana, North Carolina
- Barcodes
0082700260784200827002607859- Units affected
- 153
- About this recall
- Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.