Ortho-Clinical Diagnostics recalls VITROS Performance Verifier I
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
- Products
-
- VITROS Performance Verifier I
- Company
- Ortho-Clinical Diagnostics
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Nationwide
- Barcodes
10758750004317- Units affected
- 10,911 units
- About this recall
- VITROS Performance Verifier I