Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
Products
  • Access 2 Immunoassay Analyzer
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
15099590265380
Units affected
873 units
About this recall
Access 2 Immunoassay Analyzer, Catalog Number 81600N