Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
- Products
-
- Access 2 Immunoassay Analyzer
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
15099590265380- Units affected
- 873 units
- About this recall
- Access 2 Immunoassay Analyzer, Catalog Number 81600N