Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
Products
  • DxI 9000 Access Immunoassay Analyzer
Where it was sold

Worldwide - US Nationwide distribution in the states of FL, TX, WA.

States: Florida, Texas, Washington

Barcodes
15099590732103
Units affected
80 units
About this recall
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278