Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
- Products
-
- DxI 9000 Access Immunoassay Analyzer
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
US Nationwide. Global Distribution.
Nationwide
- Barcodes
15099590732103- Units affected
- 206 units
- About this recall
- DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.