Medline Industries, LP recalls Medline Convenience Kits: 1) ENDO KIT W/SYRINGE
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Products
-
- Medline Convenience Kits: 1) ENDO KIT W/SYRINGE
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide.
Nationwide
- Barcodes
10193489294606101934893988851019348944891710193489901399108882775380231019348972930610193489398632101934894228871088994249467510193489249460- Units affected
- 3561 units
- About this recall
- Medline Convenience Kits: 1) ENDO KIT W/SYRINGE, Model Number: DYKE1743; 2) BGMC ANGIOGRAPHY PACK-LF, Model Number: DYNJ0800934D; 3) GENERAL ENDO PACK-LF, Model Number: DYNJ0842873J; 4) OR ANGIO PACK-LF, Model Number: DYNJ43415C; 5) ANGIO PACK II, Model Number: DYNJ44066A; 6) SOUTH CATH LAB PACK, Model Number: DYNJ61978A; 7) BIOPSY ANGIO TRAY ALEX, Model Number: DYNJ63542B; 8) SURGICAL TECH KIT, Model Number: DYNJ907452; 9) ANGIOGRAPHY PACK, Model Number: DYNJC2371G; 10) SPECTRUM HEALTH CATH PACK-LF, Model Number: PHS853965G