Elekta Instrument AB recalls Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Elekta has become aware that Disposable Biopsy Needles (911933) from one batch (837838839) can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles. The sterility of the biopsy needles has not been affected. This issue has been reported from one site.
Products
  • Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System
Where it was sold

US and Australia, Austria, Belgium, France, Germany, Greece, Italy, Netherlands, Spain, Sweden, Switzerland

Units affected
601 needles
About this recall
Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, 2.1mm Diameter / 14 Gauge, containing 6 inner boxes labeled REF 911761, Disposable Biopsy Needle Kit for Leksell Stereotactic System. Each inner box of REF 911761 contains a Tyvek bag labeled REF 912465, Disposable Biopsy Needle and Aspirator Tube.