Medline Industries, LP recalls Medline Convenience Kits: 1) LITHOTOMY PACK
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Products
-
- Medline Convenience Kits: 1) LITHOTOMY PACK
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide.
Nationwide
- Barcodes
1019532724705810195327468606- Units affected
- 808 units
- About this recall
- Medline Convenience Kits: 1) LITHOTOMY PACK, Model Number: DYNJ83185; 2) LITHOTOMY PACK, Model Number: DYNJ83185A;