Baxter Healthcare Corporation recalls Hill-Rom PRO+ 36" MRS Surface
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Inconsistencies were identified with service records associated with corrections performed for a previous Medical Device Correction issued by Baxter (ref number: FA-2022-026). Your devices have been identified as potentially impacted by these service inconsistencies and therefore, Baxter will need to reassess the impacted mattresses to confirm they have been corrected as identified by the service order, and to confirm the correct serial numbers have been identified.
- Products
-
- Hill-Rom PRO+ 36" MRS Surface
- Where it was sold
-
US Nationwide distribution in the state of Colorado.
States: Colorado
- Barcodes
00887761977884- Units affected
- 232 units
- About this recall
- Hill-Rom PRO+ 36" MRS Surface, REF P7924A03, The pro+ mattress is intended for patient support and for the prevention and/or treatment of pressure injuries.