Medline Industries, LP recalls Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Products
  • Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY
Where it was sold

Worldwide distribution - US Nationwide.

Nationwide

Barcodes
10653160360421108882777337941019348919023610653160354888101934891902501019348911133010193489111019101934891999011019348919121910653160352730106531603607971019348911100210193489191257101934891903111019348911236810653160327646
Units affected
1496 units
About this recall
Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 4) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 5) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 6) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 7) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 8) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 9) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 10) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 11) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 12) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 16) CVC PACK, Model Number: CVI4500A