Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline Convenience Kits: 1) KIT ACES INSERT PEG

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Products
  • Medline Convenience Kits: 1) KIT ACES INSERT PEG
Where it was sold

Worldwide distribution - US Nationwide.

Nationwide

Barcodes
101934894790271019348982959410193489477627106531603227021065316038110510653160294313
Units affected
553 units
About this recall
Medline Convenience Kits: 1) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 2) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4; 4) NICU LINE CHANGE TRAY, Model Number: LC295; 5) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 6) PICC CATHETER INSERTION TRAY, Model Number: CVI4300