Medline Industries, LP recalls Medline Convenience Kits: 1) LVAD DRIVELINE TRAY
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Products
-
- Medline Convenience Kits: 1) LVAD DRIVELINE TRAY
- Company
- Medline Industries, LP
- Where it was sold
-
Worldwide distribution - US Nationwide.
Nationwide
- Barcodes
106531603452511065316032775200653160177473- Units affected
- 9720 units
- About this recall
- Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A