Exactech, Inc. recalls Exactech Equinoxe CAGE GLENOID
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).
- Products
-
- Exactech Equinoxe CAGE GLENOID
- Company
- Exactech, Inc.
- Where it was sold
-
Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand
Nationwide
- Barcodes
10885862201003108858622010101088586220102710885862201034- Units affected
- 4007 units
- About this recall
- Exactech Equinoxe CAGE GLENOID,POSTERIOR AUGMENT, CEMENTED, Right, Mates with: a) 38, 41, 44, 47 head, Small, Item Number 314-13-32, b) 41, 44, 47, 50, head, Medium, Item Number 314-13-33, c) 44, 47, 50, 53 head, Large, Item Number 314-13-34, d) 44, 47, 50, 53 head, Extra Large, Item Number 314-13-35; Shoulder Arthroplasty