Agfa N.V. recalls Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury
- Products
-
- Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality
- Company
- Agfa N.V.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Kingdom.
Nationwide
- Barcodes
5414904272725- Units affected
- 541 units
- About this recall