Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Agfa N.V. recalls Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury
Products
  • Agfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality
Company
Agfa N.V.
Where it was sold

Worldwide distribution - US Nationwide and the countries of Canada, Argentina, Australia, Austria, Belgium, Chile, China, Colombia, Czech Republic, Finland, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Mexico, New Zealand, Norway, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Kingdom.

Nationwide

Barcodes
5414904272725
Units affected
541 units
About this recall