Medline Industries, LP recalls Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
- Products
-
- Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Units affected
- 2020
- About this recall
- Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746