FDAMedical devicesRecalled Status: Ongoing

Checkpoint Surgical Inc recalls Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Products
  • Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead
Where it was sold

US Nationwide distribution.

Nationwide

Units affected
135
About this recall
Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.