Abbott GmbH recalls Alinity i Anti-HCV Reagent Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).
Products
  • Alinity i Anti-HCV Reagent Kit
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
00380740130657
Units affected
5110 units
About this recall
Alinity i Anti-HCV Reagent Kit, List Number 08P0521