Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips Respironics, Inc. recalls Trilogy Evo

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Potential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Products
  • Trilogy Evo
Where it was sold

Domestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, US, UY, VN, YT, ZA, ZW.

Nationwide

Barcodes
606959051980
Units affected
76,713 units
About this recall
Trilogy Evo, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients.