Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Qiagen Sciences LLC recalls EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular dia

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost
Products
  • EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular dia
Where it was sold

Nationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO, NL, NO, QA, SA, SE, SI, TH

Nationwide

Barcodes
04053228040576
Units affected
168 units: 38 units US; 130 units OUS
About this recall
EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230