Omnicell, Inc. recalls IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.
- Products
-
- IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch
- Company
- Omnicell, Inc.
- Where it was sold
-
Distributed in the US to Alabama
States: Alabama
- Units affected
- 3
- About this recall
- IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch, 5 Tyvek Pouches per shelf box, 10 shelf boxes per case box. IVX Station Reconstitution Transfer Set is used for the reconstitution of drug vials in hospital pharmacies. It is indicated for use with the IVX Station Pharmacy Compounding System.