Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 3500 Syringe pumps
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
- Products
-
- Smiths Medical Medfusion Model 3500 Syringe pumps
- Company
- Smiths Medical ASD Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
1061058604125910610586041242106105860412973061058604328810610586043291106105860430171061058603005510610586030031106105860300171061058603117510610586032318106105860317551061058603103810610586032325- Units affected
- 3821 pumps
- About this recall
- Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500