Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Model 4000 Syringe pump
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
- Products
-
- Smiths Medical Medfusion Model 4000 Syringe pump
- Company
- Smiths Medical ASD Inc.
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
106105860400781061058604005410610586040818106105860435811061058604356710610586043574150195170707501501951706962415019517070996- Units affected
- 7160 pumps
- About this recall
- Smiths Medical Medfusion Model 4000 Syringe pump, Item Numbers: a) 4000-0101-249, b) 4000-0101-50, c) 4000-0101-51, d) 4000-0105-249, e) 4000-0105-50, f) 4000-0105-51, g) 4000-0105-78, h) 4000-0106-00, i) 4000-0106-01, j) 4000-0106-231,