Smiths Medical ASD Inc. recalls Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Products
  • Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
10610586032097
Units affected
27 pumps
About this recall
Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E