Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Smiths Medical ASD Inc. recalls Smiths Medical Medfusion Guide Barrel Clamp

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Products
  • Smiths Medical Medfusion Guide Barrel Clamp
Where it was sold

Worldwide distribution.

Nationwide

Units affected
1842 guide barrel clamps
About this recall
Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716