Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
Products
  • DxI 9000 Access Immunoassay Analyzer
Where it was sold

Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.

States: New Hampshire

Barcodes
15099590732103
Units affected
17 units
About this recall
DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.