Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Datascope Corp. recalls Battery Charging Station; Model: 0998-00-0802;

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
Products
  • Battery Charging Station; Model: 0998-00-0802;
Where it was sold

Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.

States: District of Columbia, Florida, Georgia, Illinois, Indiana, Maryland, Michigan, North Carolina, Texas, Washington, Wisconsin

Barcodes
10607567111964
Units affected
28 units (20 US, 8 OUS)
About this recall
Battery Charging Station; Model: 0998-00-0802;