I.T.S. GmbH recalls I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
- Products
-
- I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1.
- Company
- I.T.S. GmbH
- Where it was sold
-
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
States: Alabama, Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, Nebraska, New Jersey, New Mexico, Nevada, Ohio, Oklahoma, Pennsylvania, South Carolina, Texas, West Virginia
- Units affected
- 510 units
- About this recall
- I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.