Route 92 Medical Inc recalls Route 92 Medical
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Their is the potential for distal tip separation at the proximal marker band of the delivery catheters
- Products
-
- Route 92 Medical
- Company
- Route 92 Medical Inc
- Where it was sold
-
US: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand
States: Alabama, Arizona, California, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, North Dakota, New Jersey, New York, Ohio, Oregon, Pennsylvania, Rhode Island, South Dakota, Tennessee, Texas, Utah, Virginia, West Virginia
- Barcodes
00853799007534- Units affected
- 200 units
- About this recall
- Route 92 Medical, Tenzing 7 Delivery Catheter, REF 1005, Rx Only, Sterile EO. Used for introduction of interventional devices into the neurovasculature.