Orthalign, Inc recalls Drill Plate
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Drill plates manufactured out of specification can result in a bone cut in excess 00800"(2.8448mm)
- Products
-
- Drill Plate
- Company
- Orthalign, Inc
- Where it was sold
-
US Nationwide distribution in the state of UT.
States: Utah
- Barcodes
00810832032434- Units affected
- 7 units
- About this recall
- Drill Plate, Ortho Dev. Trimax - Pegged, PREF Size 6-7 REF:402567 component Lantern Knee System