Abbott recalls Amplatzer TorqVue LP Delivery System (TVLP)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
- Products
-
- Amplatzer TorqVue LP Delivery System (TVLP)
- Company
- Abbott
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
0081180601156100811806011578- Units affected
- 4,850
- About this recall
- Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080