Abbott recalls Amplatzer TorqVue LP Catheter (TVLPC)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
- Products
-
- Amplatzer TorqVue LP Catheter (TVLPC)
- Company
- Abbott
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00811806012926- Units affected
- 2683
- About this recall
- Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080