Abbott recalls Amplatzer TorqVue LP Catheter (TVLPC)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
Products
  • Amplatzer TorqVue LP Catheter (TVLPC)
Company
Abbott
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00811806012926
Units affected
2683
About this recall
Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080