Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ihb Operations B.v. recalls Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Products
  • Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
Where it was sold

US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.

Nationwide

Barcodes
10841736116071
Units affected
112 units
About this recall
Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount