Ihb Operations B.v. recalls Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
- Products
-
- Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
- Company
- Ihb Operations B.v.
- Where it was sold
-
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
Nationwide
- Barcodes
10841736116071- Units affected
- 112 units
- About this recall
- Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount