Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Products
  • Beckman Coulter DxI 9000 Access Immunoassay Analyzer
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
15099590732103
Units affected
370 units
About this recall
Beckman Coulter DxI 9000 Access Immunoassay Analyzer