Electro Medical Systems SA recalls Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Produ
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).
- Products
-
- Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Produ
- Company
- Electro Medical Systems SA
- Where it was sold
-
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA REPUBLIC OF, KOSOVO, KUWAIT, LUXEMBOURG, MALAYSIA, MEXICO, MOLDOVA REPUBLIC OF, NETHERLANDS, NEW ZEALAND, OMAN, PALESTINE, PHILIPPINES, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, Thailand, UKRAINE,UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM.
States: Illinois, Texas
- Barcodes
076133531802490761335321318307613353184124- Units affected
- 54
- About this recall
- Brand Name: PIEZON Bottle Product Name: PIEZON Bottle Model/Catalog Number: EG-111 Software Version: N/A Product Description: These bottles are sold empty and are filled by the user with water - in some cases small quantities of disinfectant solutions - for dental treatments. The bottle is pressurized on the device during use. Component: YES. AIRFLOW PROPHYLAXIS MASTER and AIRFLOW PROPHYLAXIS ONE