Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls THORACIC ROBOTS

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Products
  • THORACIC ROBOTS
Where it was sold

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

States: California, Florida, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Montana, New York, Ohio, Pennsylvania, Tennessee, Texas

Barcodes
10198459269196
Units affected
14,379 kits total
About this recall
THORACIC ROBOTS, DYNJ908777B