Medline Industries, LP recalls Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
- Products
-
- Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY
- Company
- Medline Industries, LP
- Where it was sold
-
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
States: California, Florida, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Montana, New York, Ohio, Pennsylvania, Tennessee, Texas
- Barcodes
101984592735131019532764042210195327456757101984591916401019348961197710193489874594- Units affected
- 14,379 kits total
- About this recall
- Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A