Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls NAMIC Angiographic Syringe

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Products
  • NAMIC Angiographic Syringe
Where it was sold

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Nationwide

Units affected
2630369 units
About this recall
NAMIC Angiographic Syringe, labeled as follows: 1. SYRINGE,20ML,CONTROL,-,PKG, Medline SKU 70075027; 2. SYRINGE,10ML,TR/FR,RA,W/RES, Medline SKU 70085007; 3. MTO,SYRINGE,10ML,TR/FG,RA,-,PKG, Medline SKU 70085107; 4. SYRINGE,12ML,TR/FR,RA,W/RES, Medline SKU 70087007; 5. MTO,SYRINGE,12ML,TR/FG,RA,W/RES, Medline SKU 70087107; 6. SYRINGE,12ML,TR/FR,RA,W/O,RES,-,PKG, Medline SKU 70088007; 7. SYRINGE,10ML,PP/FR,RA,-,PKG, Medline SKU 70095007; 8. SYRINGE,10ML,PP/FG,RA,-,PKG, Medline SKU 70095107; 9. SYRINGE,12ML,W/RES,PP/FR,RA,PG, Medline SKU 70097007; 10. SYRINGE,12ML,PP/FG,RA,W/RES,-,PKG, Medline SKU 70097107; 11. SYRINGE,12ML,PP/FG,RA,W/O,RES, Medline SKU 70098107; 12. SYRINGE,20ML,TR/FR,RA,OEM,PG, Medline SKU 80075027; 13. SYRINGE,10ML,ANGIO,W/RA,-,OEM, Medline SKU 80085007; 14. SYRINGE,12ML,TR/FR,RA,OEM, Medline SKU 80087007; 15. SYRINGE,10ML,PP/FR,RA,-,OEM,-,PG, Medline SKU 80095007; 16. SYRINGE,10ML,PP/FG,RA,OEM,PG, Medline SKU 80095107; 17. SYRINGE,12ML,PP/FG,RA,-,OEM,-,PG, Medline SKU 80097107.