Medline Industries, LP recalls Medline medical convenience kits

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Products
  • Medline medical convenience kits
Where it was sold

Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).

Nationwide

Units affected
17902 kits
About this recall
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CATH LAB PACK-LF, Medline SKU 00-HCT055P; 2. INTERVENTIONAL CATH LAB PK-LF, Medline SKU DYNJ0376230Q; 3. ANGIOGRAPHY CATH LAB, Medline SKU DYNJ23532C; 4. ANGIO LEFT HEART PACK, Medline SKU DYNJ40632C; 5. ANGIO DRAPE PACK, Medline SKU DYNJ47952B; 6. ANGIOGRAPHIC PACK, Medline SKU DYNJ57516; 7. CATH LAB PACK, Medline SKU DYNJ58243B; 8. CATH LAB PACK, Medline SKU DYNJ65185; 9. CSU CATH MANIFOLD KIT, Medline SKU VASC1170A; 10. MANIFOLD KIT NO TRANSDUCER, Medline SKU VASC1184A.