Medical Depot Inc. recalls iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.
- Products
-
- iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to
- Company
- Medical Depot Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Albania, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, ISRAEL, Italy, JAPAN, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Maldives, Nepal, Netherlands, Norway, Oman, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Turkey, Uganda, Ukraine, United Arab Emirates.
Nationwide
- Barcodes
885304020585885304022237- Units affected
- 89,532 units
- About this recall
- iGo¿2 DC Car Adapter to be used with iGo¿2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.