Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips Ultrasound, Inc. recalls 5000 Compact Series Ultrasound Systems

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Ultrasound system with: 1) transesophageal echocardiography transducer (TEE) connected/not selected 2) 1, 2 non-TEE connected 3) Imaging with non-TEE 4) Selected/not selected imaging transducer removed 5) Transducer connected 6) Non-TEE imaging continues, Then though TEE is unselected, it may have power, no temperature monitoring/control, and may result in temperature above limit and patient burns
Products
  • 5000 Compact Series Ultrasound Systems
Where it was sold

Worldwide - US Nationwide distribution including in the states of CA, CO, CT, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, NC, NH, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA and the countries of Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Lithuania, Netherlands, New Zealand, Oman, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates.

Nationwide

Units affected
363
About this recall
5000 Compact Series Ultrasound Systems, Product/REF: 5500 CV/795141, 5500 P/795143, 5500 W/795144, 5500 G/795145, and Product/REF: 5300 W/795146, 5300 P/795147, 5300 G/795148, used with Deluxe and Premium Carts that come with multiport adapters.