Randox Laboratories Ltd. recalls Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug mole

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
Foam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
Products
  • Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug mole
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
05055273214970
Units affected
40 units
About this recall
Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115