B Braun Medical Inc recalls B Braun Hemodialysis Bloodlines
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
- Products
-
- B Braun Hemodialysis Bloodlines
- Company
- B Braun Medical Inc
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of Canada.
Nationwide
- Units affected
- 4,884 units
- About this recall