FDAMedical devicesRecalled Status: Ongoing

Karl Storz Endoscopy recalls LASER Application Instrument

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.
Products
  • LASER Application Instrument
Where it was sold

Worldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX, KY, WI, MI, MN, AR, NE, IL, NY, HI and the country of Argentina.

States: Arkansas, California, Florida, Hawaii, Illinois, Kentucky, Massachusetts, Michigan, Minnesota, Nebraska, New York, Ohio, Texas, Wisconsin

Units affected
85
About this recall
LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000