Karl Storz Endoscopy recalls LASER Application Instrument
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Laser surgical instruments have no specific evidence of a validated reprocessing method, therefore these instruments lack sterility assurance, so once the instruments have gone through reprocessing after use and then are used again, there is an increased risk of the patient being exposed to infection.
- Products
-
- LASER Application Instrument
- Company
- Karl Storz Endoscopy
- Where it was sold
-
Worldwide distribution - US Nationwide in the states of OH, CA, FL, MA, TX, KY, WI, MI, MN, AR, NE, IL, NY, HI and the country of Argentina.
States: Arkansas, California, Florida, Hawaii, Illinois, Kentucky, Massachusetts, Michigan, Minnesota, Nebraska, New York, Ohio, Texas, Wisconsin
- Units affected
- 85
- About this recall
- LASER Application Instrument, 23 cm, REF: 8694; LEUNIG-GREVERS LASER Application, REF: 461000