Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beta Bionics, Inc. recalls iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
Products
  • iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
Where it was sold

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, LA, MA, MD, ME, MI, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WY AND WASHINGTON DC.

Nationwide

Units affected
15 units
About this recall
iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software